Remote GPIB Interface
78
Serial BERT 12.5 Gb/s User Guide
Command
:TRIGger:SAMPle
Type
Command
Description
Takes sample of current measurement. Does not return any values, but allows
the values to be queried by other commands.
Example
See below
Command
:SENSe:DATA? [BCOunt | ECOunt | ETIMe | BER | ALL]
Type
Querry
Description
Returns the selected parameter from the current instrument sample, or returns
all parameters as a vector if ALL or no parameter is specified.
BCOunt Total number of bits counted
ECOunt Total number of errors counted
ETIMe Elapsed time (seconds)
BER Calculated Bit Error Rate
The vector is as follows: BCOunt,ECOunt,ETIMe,BER
Parameters
[BCOunt | ECOunt | ETIMe | BER | ALL]
Note
Need to run ":TRIG:INIT" and ":TRIG:SAMPLE” first.
Result
Described above