Operation
Serial BERT 12.5 Gb/s User Guide 45
Text
Type
Description
Default
ε time (sec)
Displays the elapsed time of the BER
measurement, in seconds. If the error
detector is on, the el
apsed time will
increment in real time. If the error detector
is off, the elapsed time shown is from the
previous session.
Ampl (V)
Displays the single
-ended PRBS output
amplitude, in volts. This voltage is per side,
multiply by 2 when the output is used
d
ifferentially.
Freq (GHz)
Displays the internal clock frequency, in
GHz.
When using an external clock (Synth
= 0), then the Freq adjust range is
increased to 500 MHz min, 12.5 GHz max.
The N4962A Frequency setting should be
set to the frequency of the ext
ernal clock
so that the N4962A frequency dependent
parameters are correctly set.
Ø
Displays the error detector electronic clock
phase shift setting, in degrees. The phase
sh
ift amount ranges from 0 to +358
degrees (one unit interval).
When Ø is shown on t
he display, pressing
the Adjust
->Config State PLUS button will
prompt the N4962A to automatically
determine the optimum detector clock
phase. This is detailed in Section
4.6.
Config State
Displays one of
ten configuration states
that can be modified with the Adjust
-
>Config State PLUS or MINUS buttons.
Different state can be selected with the
Scroll UP or DOWN buttons, but the light