Operation
34
Serial BERT 12.5 Gb/s User Guide
4.1.1 Performance Recommendations
Follow the following recommendations for best performance:
When using differential mode connection for IN/IN¯¯, OUT/OUT¯¯¯, or
DataThru/DataThru¯¯¯¯¯¯ , ensure the cables are phase balanced. If the
electrical length of one cable is a significant fraction of a unit interval
longer than the other, the quality of the differential signal will be
degraded and BER will increase.
Keep cable lengths short and minimize number of cable bends.
Terminate all unused RF connectors with 50 Ω terminations.
4.1.2 Connector Care
The N4962A serial BERT 12.5 Gb/s features high-quality SMA connectors, and
two 2.92 mm K-connectors for the front-panel OUT/OUT¯¯¯ interface. Agilent
Technologies does not recommend using male SMA connectors with the
female K-connectors; the male SMA pin has looser tolerances than the female
K sleeve, and may damage the connector. Connector damage will degrade
signal fidelity.
Refer to the N4960-90030 N495xA through N498xA Connector Care Reference
Guide at www.agilent.com/find/N4962A
.
Agilent Technologies also recommends the following:
Use a 7-10 in-lbs torque wrench when attaching connectors.
Consider using connector savers to prolong performance and minimize
damage.
Differential connectors may be used single-ended if second end
terminated in 50 Ω.
Terminate all unused RF connectors with 50 Ω terminations.