Agilent Technologies N4962A Guia do Utilizador Página 73

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 88
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 72
Serial BERT 12.5 Gb/s User Guide
73
Command
:SENSe:APTime {1 <= value <= 5000}
Type
Command
Description
Sets the desired autophase measurement time, in 1/10ths of a millisecond.
The lowest number is 1 (0.1 ms). The highest number is 5000 (500 ms).
Parameters
{1 <= value <= 5000}
Example
OUTPUT 725; ":SENS:APT 300"
OUTPUT 725; ":SENS:APT?"
ENTER 725; value$
PRINT value$ ! should be “300”
Command
:SENSe:ROSCillator:PHASe?
Type
Query
Description
Returns the error detector clock phase.
Phase is selected in 2 degree increments for data rates greater then 5 GHz to
12.5 GHz. Data rates between 500 MHz and less than 5 GHz offer values of 0,
90, 180, 270 degrees only.
Results
{0 <= value <= 358}
Example
See below