8 Frequency Sweep Commands
90 U2761A Programmer’s Reference
[SOURce:]FREQuency:STARt
Syntax
[SOURce:]FREQuency:STARt <frequency>
This command sets the start frequency.
[SOURce:]FREQuency:STARt?
This query returns the start frequency in Hz.
Parameter
Examples
The following command sets the start frequency to 500 Hz.
FREQ:STAR 500
The following query returns the start frequency.
FREQ:STAR?
Typical response:
+5.0000000000000E+02
Item Type Range of values Default value
frequency Numeric 1 μHz to 20 MHz (for sine and square)
1 μHz to 200 kHz (for ramp and arbitrary)
100 Hz
SOUR : FREQ space
?
frequency:STAR