6 Modulation Commands
78 U2761A Programmer’s Reference
[SOURce:]PM:STATe
Syntax
[SOURce:]PM:STATe {0|OFF|1|ON}
This command disables or enables the phase modulation.
[SOURce:]PM:STATe?
This query returns the phase modulation state as 0 if the modulation is
disabled, and 1 if the modulation is enabled.
Parameter
Remark
The U2761A allows only one modulation mode to be enabled at a time,
and does not allow modulation when sweep is enabled.
Examples
The following command enables the phase modulation.
PM:STAT ON
The following query returns the modulation state.
PM:STAT?
Typical response:
1
Item Type Range of values Default value
state Boolean 0|OFF|1|ON 0
SOUR : PM space
?
0|OFF
1|ON
:STAT