7 Shift Keying Commands
80 U2761A Programmer’s Reference
[SOURce:]ASKey:INTernal:RATE
Syntax
[SOURce:]ASKey:INTernal:RATE <rate in Hz>
This command sets the rate at which the output amplitude “shifts
between two preset amplitudes.
[SOURce:]ASKey:INTernal:RATE?
This query returns the amplitude-shift keying (ASK) rate in Hz.
Parameter
Remark
The modulating waveform is a Square wave with a 50% duty cycle.
Examples
The following commands sets the ASK rate to 10 Hz.
ASK:INT:RATE 10
The following query returns the ASK rate.
ASK:INT:RATE?
Typical response:
+1.0000000000000E+01
Item Type Range of values Default value
rate Numeric 2 mHz to 100 kHz 10 Hz
SOUR : ASK space
?
rate in Hz:INT :RATE
Vista de página 87