Modulation Commands 6
U2761A Programmer’s Reference 77
[SOURce:]PM:DEViation
Syntax
[SOURce:]PM:DEViation <deviation in degrees>
This command sets the phase deviation in degrees. The deviation setting
represents the peak variation in phase of the modulated waveform from
the carrier waveform.
[SOURce:]PM:DEViation?
This query returns the phase deviation in degrees.
Parameter
Examples
The following command sets the phase deviation to 10 °.
PM:DEV 10
The following query returns the phase deviation.
PM:DEV?
Typical response:
+1.0000000000000E+01
Item Type Range of values Default value
deviation Numeric 0 ° to 360 ° 180 °
SOUR : PM space
?
deviation in degrees:DEV