4 Output Configuration Commands
50 U2761A Programmer’s Reference
[SOURce:]VOLTage[:LEVel][:IMMediate]:UNIT
Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate]:UNIT {VPP|VRMS|DBM}
This command sets the unit for the output amplitude.
[SOURce:]VOLTage[:LEVel][:IMMediate]:UNIT?
This query returns the unit for the output amplitude as VPP, VRMS, or
DBM.
Parameter
Examples
The following command sets the unit of the output amplitude to Vpp.
VOLT:UNIT VPP
The following query returns the output amplitude unit.
VOLT:UNIT?
Typical response:
VPP
Item Type Range of values Default value
unit Discrete VPP|VRMS|DBM VPP
SOUR : VOLT space
?
:LEV :IMM :UNIT VPP
VRMS
DBM