Agilent Technologies AGILENT U2761A Manual de Serviço Página 143

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 169
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 142
CALibration Commands 14
U2761A Programmer’s Reference 135
CALibration:STRing
Syntax
CALibration:STRing <quoted string>
This command stores a message in nonvolatile calibration memory. Storing
a message will overwrite any message previously stored in memory.
CALibration:STRing?
This query reads the calibration message and returns a quoted string.
Parameter
Examples
The following command stores the calibration message.
CAL:STR '01 JAN 2008:CALIBRATED'
The following query returns the quoted string.
CAL:STR?
Typical response:
''01 JAN 2008:CALIBRATED''
Item Type Range of values Default value
string String Up to 40 alphanumeric characters N/A
CAL space
?
quoted string:STR
Vista de página 142
1 2 ... 138 139 140 141 142 143 144 145 146 147 148 ... 168 169

Comentários a estes Manuais

Sem comentários