5 Pulse Configuration Commands
62 U2761A Programmer’s Reference
[SOURce:]FUNCtion:PULSe:DCYCle
Syntax
[SOURce:]FUNCtion:PULSe:DCYCle <percent>
This command sets the pulse duty cycle percentage.
[SOURce:]FUNCtion:PULSe:DCYCle?
This query returns the current pulse duty cycle in %.
Parameter
Examples
The following command sets the pulse duty cycle to 50%.
FUNC:PULS:DCYC 50
The following query returns the pulse duty cycle.
FUNC:PULS:DCYC?
Typical response:
+5.0000000000000E+01
Item Type Range of values Default value
percent Numeric 0% to 100%
[1]
[1] The pulse duty cycle is limited by minimum pulse width restrictions which prevent you from setting
exactly 0% or 100%. Refer to the U2761A USB Modular Function/Arbitrary Waveform Generator
User’s Guide for more information.
50%
SOUR : FUNC space
?
:PULS :DCYC percent