Agilent Technologies AGILENT U2761A Manual de Serviço Página 133

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 169
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 132
Status Reporting Commands 13
U2761A Programmer’s Reference 125
STATus:QUEStionable:CONDition?
Syntax
STATus:QUEStionable:CONDition?
This query returns the binary-weighted sum of all bits set in the
Questionable Data condition register. The condition register is a read-only
register, which holds the live (unlatched) questionable status of the
instrument. The bits are not cleared when you read the register.
Remarks
For more information on the Questionable condition register, refer to
Chapter 1, “Status system diagram” on page 12. The bit definitions for the
Questionable Data register are listed in “Questionable Data register” on
page 13.
Example
The following query reads the Questionable condition register.
STAT:QUES:COND?
Typical response:
+544
STAT ?:QUES :COND
Vista de página 132