© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 4
Introduction
Tools for removing the handler:
■ 2.5-millimeter hex wrench
■ 5-millimeter hex wrench
■ EFS removal cart and blower
NOTE
For instructions on removing the board handler,
refer to the procedure in the 44990A EFS Board
Handler Manual (44990-90004), Chapter 5. Save
all hardware.
4 Run Diagnostics, especially the DUT power supply
tests with a pin verification fixture. Correct any
problems detected before installing the DUT supply.
5 Turn power to the testhead OFF using these steps:
Execute testhead power off and switch the system
PDU(s) OFF.
CAUTION
✸
Before beginning these installation procedures,
verify that power to the testhead and support bay
is turned OFF. Execute a testhead power off
and switch the system PDU(s) OFF.
6 For 307X systems, it is not necessary but may be
helpful to remove the support bay covers as
described below (save all hardware):
a Using a cabinet key, open the front and rear doors.
You may remove the doors, but it’s not necessary.
b Using a #2 Pozidriv screwdriver, first remove the
top cover; it is held by two screws in the rear.
Then remove the left and right side covers; they
are each held by two screws at the top.
7 For 317X systems, remove the testhead covers (save
all hardware):
a Rotate the testhead to the 0-degree (shipping)
position.
b If the probe is in the probe holder on the safety
skirt, move it to a safe location.
c Using a 5-millimeter hex wrench, remove the
safety skirt.
d Using a T20 Torx driver, remove the top and
bottom covers; they are each held by one screw on
the front of the testhead.
e Rotate the testhead to the 90-degree (service)
position.
f Using a 1/4-inch flat blade screwdriver, remove
the rear cover.