© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 56
Connecting the DUT Power Supply Cables
6642A (E3784A) Supply
The 6642A DUT Power Supply is a one-output,
high-current supply. It can only be connected to
channels 5 and 6 on the ASRU Card. Two ASRU
channels are required to handle the high current.
Connect the supply to the highest numbered module
with channel 5 and 6 available. Use cable E4000-61602
for a 307X system or E4000-61606 for a 317X or 327X
system. The spade lugs for the sense lines must be
removed, the wires stripped and attached to the
connector shipped with the power supply as shown in
Figure 26.
Figure 26 Sense line connections for the 6642A DUT power supply
connect.wpg
WHT and WHT
GRN and BLU
Channel 5
Channel 6