1100 LC Modules Control Manual - 47 - 03-914947-30:Rev.6
Others Parameter Group
In the Needle Offset field, enter the vertical needle offset value.
In the Sample Flush-Out factor field, enter the flush out factor
value that specifies the volume of mobile phase pumped through
the injector after each injection.
Check Switch valve to bypass after sample flush-out to leave the
valve in the bypass position even past analysis.
In the Wait time field, enter the wait time after drawing value.
Select the injection mode in the scrolling list. Three choices are
available: Standard injection, Injection with wash and Injection
user program.
• Standard injection: only the default screen is
displayed. No other parameters are displayed.
• Injection with wash: when selecting this injection
mode, another tab is displayed allowing the user to
define multiple parameters.
Select in the drop down list the type wash. Three
choices are available: Wash needle in flush port,
Wash needle in wash vial, and Flush the seat.
Depending of the wash type, the following fields will
be available.
Rack: define in which rack the wash vial is located.
Vial: define the location of the wash vial.
Cycles: define the number of time the wash
procedure will be done.
Delay: define the time the needle will stay in the
wash port or flush the seat.