1100 LC Modules Control Manual - 155 - 03-914947-30:Rev.6
Each counter modification is automatically listed in the
maintenance audit trail.
Press
to add the current counters table in
the maintenance audit trail.
Counters definition
Volume pumped A: total pumped volume of solvent A.
Volume pumped B: total pumped volume of solvent B.
Volume pumped C: total pumped volume of solvent C.
Volume pumped D: total pumped volume of solvent D.
Total Volume pumped: Sum of the pumped volume of each
solvent channel.
Pumped volume: Total pumped volume. It is incremented by the
stroke volume with each piston reversal.
Wear: The seal wear value is derived from system pressure and
flow, is used as an indicator of seal wear in the pump head. The
value increments continually with pump usage.
Inlet valve switches: total number of active-inlet valve switches.
The value increments by one with each pump cycle.
Outlet valve switches: total number of outlet-valve switches. The
value increments by one with each pump cycle.
Each take in phase: total number of times the plungers reached
the take-in phase state.
Each deliver phase: total number of times the plungers reached
the delivery phase state.
Solvent valve A1/A2: Total number of solvent selection valve A
switches.
Solvent valve B1/B2: Total number of solvent selection valve B
switches.
MCGV switches: Total number of interrupts for MCGV switches.