Agilent Technologies 1100 Series Guia de Instalação Página 150

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 184
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 149
1100 LC Modules Control Manual - 150 - 03-914947-30:Rev.6
Press to close the gripper and move the fingers
as close as possible to the top of the vial.
Use the arrows keys for correcting rotational, horizontal and
vertical position.
Press
to open the gripper and move it in such a
way that the cap of the vial and rubber of gripper fingers have
the same height.
Press
to validate the new vial 1 position.
Press
and repeat the same steps to correct the vial
81 position. Once the vial 81 is aligned, press
.
Once vial 1 and 81 positions are corrected, press the
button to realign the tray. The new alignment
is stored permanently and the module is reset.
Press
during the procedure to cancel all
correction and set previous alignment.
Vista de página 149
1 2 ... 145 146 147 148 149 150 151 152 153 154 155 ... 183 184

Comentários a estes Manuais

Sem comentários