CONFigure
The CONFigure command subsystem configures the multimeter to perform the
specified measurement with the given range and resolution.
CONFigure does not
make the measurement after setting the configuration. Use the
INITiate command to
place the multimeter in the wait-for-trigger state and store readings in mainframe or
command module memory. Or, use the
READ? command to make the measurement
and send the readings to the output buffer when the trigger is received.
Executing
CONFigure is equivalent to configuring the multimeter with the low-level
commands shown in the following table.
Command Setting
VOLTage:RANGe
RESistance:RANGe
As specified, or autorange.
VOLTage:RESolution
RESistance:RESolution
As specified, or as a function of range, integration time,
or aperture time.
VOLTage:APERture
RESistance:APERture
16.7 ms (60 Hz) or 20 ms (50 Hz), or based on specified
resolution.
VOLTage:NPLC
RESistance:NPLC
1 PLC, or based on specified resolution.
CALibration:ZERO:AUTO ON (autozero is performed after every measurement).
RESistance:OCOMpensated OFF (applies to resistance measurements only).
TRIGger:SOURce
TRIGger:COUNt
TRIGger:DELay
IMM (trigger signal is always true).
1
AUTO (DC volts/resistance: 0 s, AC volts: 0.5 s).
SAMPle:COUNt
SAMPle:SOURce
1
IMM
Subsystem Syntax CONFigure
:FRESistance [<range>[,<resolution>]] [,<channel_list>]
:RESistance [<range>[,<resolution>]] ,<channel_list>
:TEMPerature <transducer>,<type> ,<channel_list>
:VOLTage:AC [<range> [,<resolution>]] [,<channel_list>]
:VOLTage[:DC] [<range> [,<resolution>]] [,<channel_list>]
NOTE: If range and resolution are not specified (that is, if you use the default
values), use a space rather than a comma before the channel_list parameter.