Agilent Technologies E1326B Manual do Utilizador Página 21

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 291
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 20
Setting the Logical
Address Switch
Figure 2-1 shows the location and settings of the multimeter’s logical
address switch.
The switch has a factory setting of 24 which is equivalent to a secondary
HP-IB address of 03. If you have more than one multimeter, you must change
the logical address to some other multiple of 8 (for example, 32, 40, 48...), as
there can only be one instrument per secondary address.
Figure 2-1. HP E1326/1411 Logical Address Switch Settings
22 Configuring the HP E1326B/E1411B Multimeter Chapter 2
Vista de página 20
1 2 ... 16 17 18 19 20 21 22 23 24 25 26 ... 290 291

Comentários a estes Manuais

Sem comentários