Agilent Technologies E1326B Manual do Utilizador Página 29

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 291
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 28
Connecting
Multiplexers
In a scanning multimeter configuration, the multimeter is connected to the
multiplexers with an analog bus cable, or with the analog bus cable and a
digital bus cable. The cable(s) used is determined as follows:
1. If the scanning multimeter uses relay multiplexers only, the analog
bus cable is used.
2. If the scanning multimeter uses FET multiplexers only, the analog
bus cable and the digital bus cable are used.
3. If the scanning multimeter uses a combination of relay and FET
multiplexers, only the analog bus cable is used.
Figure 2-8 shows how the analog and digital bus cables are connected.
Figure 2-8. Connecting the Analog and Digital Bus Cables
30 Configuring the HP E1326B/E1411B Multimeter Chapter 2
Vista de página 28
1 2 ... 24 25 26 27 28 29 30 31 32 33 34 ... 290 291

Comentários a estes Manuais

Sem comentários