Example: Changing
Measurement
Functions with
FUNCtion
10 !Configure for DC voltage measurement.
20 CONF:VOLT:DC
30 !
Put multimeter in wait-for-trigger state, take reading.
40 READ?
50 !
Enter reading into computer.
60 !Change function to 4-wire resistance.
70 FUNC:FRES
80 !
Put multimeter in wait-for-trigger state, take reading.
90 READ?
100 !
Enter reading into computer.
In addition to the function change the range, resolution, aperture time, and
integration time for the second measurement are set to either their reset or
last programmed values. The triggering parameters remain as set by
CONFigure.
Multimeter Parameters
Many of the parameters set by MEASure, CONFigure, and low-level
commands configure the multimeter’s analog-to-digital (A/D) converter and
other portions of its measurement circuitry. These parameters include:
range
resolution
aperture and integration time
autozero
offset compensation
This section describes these parameters. The settings are summarized in
Table 4-5 on page 92.
Chapter 4 Understanding the HP E1326B/E1411B Multimeter 91