RESistance:RANGe?
[SENSe:]RESistance:RANGe? [MIN | MAX] returns one of the following numbers
to the output buffer:
The present resistance range is selected if
MIN or MAX is not specified. Only the ranges
available with the
RANGe command are returned. For example, if CONFigure selects
the 232 range,
256 is the range returned.
The minimum resistance range available (256 ) if
MIN is specified.
The maximum resistance range available (1048576 ) if
MAX is specified.
Example Querying the Measurement Range
RES:RANG 256
!Select 256
range.
RES:RANG? !Query multimeter to return the present
range.
enter statement !Enter value into computer.
RESistance:RANGe :AUTO
[SENSe:]RESistance:RANGe:AUTO <mode> enables or disables the autorange
function for resistance measurements.
Parameters
Parameter Name Parameter Type Range of Values Default Units
<mode>
boolean OFF | 0 | ON | 1 none
Comments You can substitute decimal values for the OFF ("0") and ON ("1") parameters.
When autoranging is ON, the multimeter samples the input before each
measurement and selects the appropriate range.
If you explicitly select a range using RESistance:RANGe, autoranging is turned OFF.
In order to specify an aperture time of 10 µs, you must select a fixed range
(
RESistance:RANGe:AUTO OFF, for example).
Related Commands: CONFigure, RESistance:RANGe
*RST Condition: RES:RANG:AUTO ON
Example Disabling Autoranging
RES:RANG:AUTO OFF !Disable autorange.
167 Command Reference Chapter 5