MEMory
The MEMory command subsystem enables you to store multimeter readings in
shared memory (an external VME memory card).
Subsystem Syntax MEMory
:VME:ADDRess <address>
:VME:ADDRess? [MIN | MAX]
:VME:SIZE <bytes>
:VME:SIZE? [MIN | MAX]
:VME:STATe <mode>
:VME:STATe?
MEMory
SubsystemData
Format
The multimeter sends readings to an external VME memory card in IEEE-754
32-bit notation (this is the IEEE standard for binary floating-point representation).
:VME:ADDRess
MEMory:VME:ADDRess <address> sets the address of the external memory board
in A24 memory address space.
Parameters
Parameter Name Parameter Type Range of Values Default Units
<address>
numeric 2097152 - 14680060
#H200000 - #HDFFFFC
none
Comments You can specify the address location in decimal or hexadecimal (#H....).
MIN
sets the address to 2097152 (#H200000). MAX sets the address to 14680060
(#HDFFFFC) - to store one reading.
The VME address specified is based on the memory card configuration. Refer to the
memory card manual for configuration information.
*RST Condition: MEM:VME:ADDR #H200000
Example Setting the VME Memory Address
MEM:VME:ADDR #H800000 !Set memory address location.
:VME:ADDRess?
MEMory:VME:ADDRess? [MIN | MAX] returns one of the following numbers to
the output buffer:
The present decimal address selected if
MINimum or MAXimum are not specified.