Register Offset The register offset is the register’s location in the block of 64 address bytes.
For example, the multimeter’s Command Register has an offset of 08
16
.
When you write a command to this register, the offset is added to the base
address to form the register address:
C600
16
+ 08
16
= C608
16
1FC600
16
+ 08
16
= 1FC608
16
or
50,688 + 8 = 50,696
2,082,304 + 8 = 2,082,312
Accessing the
Registers
Table C-1 summarizes several of the programming methods used to access
the E1326B/E1411B multimeter registers.
Table C-1. Accessing the Multimeter Registers
Computer Programming Method Base Address
E1300/01 IBASIC
(Absolute Addressing)
(Select Code 8)
READIO (-9826, Base_addr + offset)
WRITEIO -9826, Base_addr + offset; data
(positive select code = byte read or write
negative select code = word read or write)
READIO (8, Base_addr + reg number)
WRITEIO 8, Base_addr + reg number; data
Base_addr = 1FC000
16
+ (LADDR * 64)
16
or
= 2,080,768 + (LADDR * 64)
offset = register offset (Figure C-2)
Base_addr = LADDR * 256
reg number = register offset (Figure C-2)/2
External Computer
(over HP-IB to
E1300A/E1301A
Mainframe or E1406A
Command Module)
VXI:READ? logical_address, offset
VXI:WRITE logical_address, offset, data
DIAG:PEEK? Base_addr + offset, width
DIAG:POKE Base_addr + offset, width, data
module logical address setting (LADDR)
offset = register offset (Figure C-2)
Base_addr = 1FC000
16
+ (LADDR * 64)
16
or
= 2,080,768 + (LADDR * 64)
offset = register offset (Figure C-2)
V/382 Embedded
Computer
(C-Size system)
READIO (-16, Base_addr + offset)
WRITEIO -16, Base_addr + offset; data
(positive select code = byte read or write
negative select code = word read or write)
Base_addr = C000
16
+ (LADDR * 64)
16
or
= 49,152 + (LADDR * 64)
offset = register offset (Figure C-1)
HP RADI-EPC7
Embedded Computer
with SICL
iwpeek((unsigned short *)(base_addr + offset))
iwpoke((unsigned short *)
(base_addr + offset),data)
INST device_name;
device_name = iopen (“vxi, logical address”);
base_addr = imap
(device_name, I_MAP_VXIDEV,0,1,NULL);
LADDR: E1326B/E1411B logical address = 24
(LADDR * 64)
16
: Multiply quantity then convert to a hexadecimal number (e.g. (24 * 64)
16
= 600
16
)
When using DIAG:PEEK? and DIAG:POKE, the width (number of bits) is either 8 or 16.
202 HP E1326B/E1411B Multimeter Register-Based Programming Appendix C