Agilent Technologies E1326B Manual do Utilizador Página 214

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 291
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 213
Querying
Parameters
Multimeter parameters are queried as shown in Figure C-7.
Comments
The registers used are:
Status Register (base + 04
16
)
Command Register (base + 08
16
)
Query Response Register (base + 08
16
)
Status bit 7 is monitored to determine when a query opcode
(command) can be written to the Command Register.
Status bit 1 is monitored to determine when the response to the query
is in the Query Response Register.
Figure C-7. Querying Multimeter Parameters
Appendix C HP E1326B/E1411B Multimeter Register-Based Programming 215
Vista de página 213
1 2 ... 209 210 211 212 213 214 215 216 217 218 219 ... 290 291

Comentários a estes Manuais

Sem comentários