Agilent Technologies InfiniiVision 5000 Series Guia do Utilizador Página 286

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 296
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 285
286 5000 Series Oscilloscope User’s Guide
8 Power and Environmental Conditions
Transient Withstand Capability
CAUTION
Maximum input voltage for analog inputs
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 100 Vrms, 400 Vpk
with N2863A 10:1 probe: CAT I 600 V, CAT II 300 V (DC + peak AC)
with 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
with 50 input: 5 Vrms
CAUTION
Do not exceed 5 Vrms in 50 mode on the 2-channel models. Input
protection is enabled in 50 mode, and the 50 load will disconnect
if greater than 5 Vrms is detected. However, the input could still be
damaged, depending on the time constant of the signal. The 50 input
protection mode only functions when the oscilloscope is powered on.
Vista de página 285
1 2 ... 281 282 283 284 285 286 287 288 289 290 291 ... 295 296

Comentários a estes Manuais

Sem comentários