Agilent Technologies InfiniiVision 5000 Series Guia do Utilizador Página 112

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 296
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 111
112 5000 Series Oscilloscope User’s Guide
3 Triggering the Oscilloscope
To use I
2
C triggering
An I
2
C (Inter-IC bus) trigger setup consists of connecting the
oscilloscope to the serial data (SDA) line and the serial clock
(SCL) line, then triggering on a stop/start condition, a restart, a
missing acknowledge, an EEPROM data read, or on a read/write
frame with a specific device address and data value.
1 Press the More key in the Trigger section of the front panel,
rotate the Entry knob until I
2
C is displayed in the Trigger
softkey, then press the Settings softkey to display the I
2
C
trigger menu.
NOTE
For I
2
C decode setup see page 217.
Signals
menu
Trigger on:
condition
Address
Data 2
value
Trigger level
or threshold
Currently selected Clock
or Data channel
I
2
C trigger
Return to
previous menu
Data
value
Vista de página 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 295 296

Comentários a estes Manuais

Sem comentários