Agilent Technologies InfiniiVision 5000 Series Guia do Utilizador Página 109

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 296
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 108
Triggering the Oscilloscope 3
5000 Series Oscilloscope User’s Guide 109
softkey, then press the Settings softkey to display CAN trigger
menu.
2 Repeatedly press and release the Trigger: softkey or rotate the
Entry knob to select the trigger condition.
SOF - Start of Frame — The oscilloscope triggers at the start
of a frame.
Remote Frame ID (RTR) — The oscilloscope triggers on
remote frames with the specified ID. Press the Bits softkey
to select the ID. For details about using the Bits softkeys,
press and hold the softkey in question to display the
built-in help for this subject.
Data Frame ID (~RTR) - The oscilloscope will trigger on data
frames matching the specified ID. Press the Bits softkey to
select the ID.
Remote or Data Frame ID - The oscilloscope will trigger on
remote or data frames matching the specified ID. Press the
Bits softkey to select the ID.
Data Frame ID and Data - The oscilloscope will trigger on
data frames matching the specified ID and data. Press the
Bits softkey to select the ID and set up the number of data
bytes and values.
Trigger
condition
Signals
Menu
Trigger level
or threshold
Currently selected
signal source
CAN trigger
Return to
previous menu
Bits
Selector
Vista de página 108
1 2 ... 104 105 106 107 108 109 110 111 112 113 114 ... 295 296

Comentários a estes Manuais

Sem comentários