Displaying Data 5
5000 Series Oscilloscope User’s Guide 229
Interpreting Decoded CAN Data
Frame ID appears in hex digits in yellow. Frames of 11 or 29
bits are automatically detected.
Remote frame (RMT) appears in green.
Data length code (DLC) appears in blue.
Data bytes appear in hex digits in white for data frames.
Cyclic redundancy check (CRC) appears in hex digits in blue
when valid, or in red to indicate that the oscilloscope’s
hardware decode calculated a CRC that is different from the
incoming CRC data stream.
White lines show an idle bus.
Blue rectangles show an active bus (inside a packet/frame).
Decoded hexadecimal data values may extend beyond the
frame if the timebase setting is too slow.