118 5000 Series Oscilloscope User’s Guide
3 Triggering the Oscilloscope
To use LIN triggering
LIN (Local Interconnect Network) trigger setup consists of
connecting the oscilloscope to a serial LIN signal.
LIN triggering will trigger on the rising edge at the Sync Break
exit of the LIN single-wire bus signal that marks the beginning
of the message frame. If the N5424A CAN/LIN Automotive
Triggering and Decode option is installed on your oscilloscope,
the Frame ID trigger type will also be available.
A LIN signal message frame is shown below:
NOTE
For LIN decode information see page 233.
Sync Break Exit
Sync
Break
Sync
Field
Identifier
Break
Data
Fields
Checksum
Field