Agilent Technologies InfiniiVision 5000 Series Guia do Utilizador Página 137

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 296
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 136
Triggering the Oscilloscope 3
5000 Series Oscilloscope User’s Guide 137
To trigger on a specific field of the video signal
To examine the components of a video signal, trigger on either
Field 1 or Field 2 (available for interleaved standards). When a
specific field is selected, the oscilloscope triggers on the rising
edge of the first serration pulse in the vertical sync interval in
the specified field (1 or 2).
1 Press the Trigger More key, then press the TV softkey.
2 Press the Settings softkey, then press the Standard softkey to
select the appropriate TV standard.
3 Press the Mode softkey and select Field1 or Field2.
Figure 10 Triggering on Field 1
Vista de página 136
1 2 ... 132 133 134 135 136 137 138 139 140 141 142 ... 295 296

Comentários a estes Manuais

Sem comentários