Agilent Technologies J-BERT N4903B Manual de Serviço Página 359

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 406
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 358
MEAS:JTOL:CHAR(*):PAR:SEARch:DLIN:STEP?
The command sets the jitter amplitude step size for a downwards linear search.
The unit is UI.
The query returns the present step size.
MEAS:JTOL:CHAR(*):PAR:SEARch:ULOG:STARt[?]
IAgilentN490xJCharacterizationMeasurement.SetSearchParameter/
GetSearchParameter
MEAS:JTOLerance:CHAR(*):PARameter:SEARch:ULOG:STARt <NR3>
MEAS:JTOL:CHAR(*):PAR:SEARch:ULOG:STARt?
The command sets the jitter amplitude start value for an upwards logarithmic
search. The unit is UI.
The query returns the present value.
MEAS:JTOL:CHAR(*):PAR:SEARch:ULOG:RATIo[?]
IAgilentN490xJCharacterizationMeasurement.SetSearchParameter/
GetSearchParameter
MEAS:JTOLerance:CHAR(*):PARameter:SEARch:ULOG:RATIo <NR3>
MEAS:JTOL:CHAR(*):PAR:SEARch:ULOG:RATIo?
The command sets the ratio between two successive jitter amplitudes for an
upwards logarithmic search. The ratio must be greater than 1.
For example, if the start value is 0.1 UI and the ratio is 2, the jitter amplitude from
one measured point to the next will be 0.1 UI, 0.2 UI, 0.4 UI, 0.8 UI, 1.6 UI, and so
on.
The query returns the present ratio.
MEAS:JTOL:CHAR(*):PAR:SEARch:DLOG:STOP[?]
IAgilentN490xJCharacterizationMeasurement.SetSearchParameter/
GetSearchParameter
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 359
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Vista de página 358