The command measures the DC level at the CDR input and sets the measured value
as CDR threshold.
The query returns the current DC level at the CDR input.
SENSe2:VOLTage:EDGe[?]
IAgilentN490xEDClockIn.ActiveEdge (IVI-compliant)
SENSe2:VOLTage:EDGe NEGative | POSitive
SENSe2:VOLTage:EDGe?
Sets the active edge of the clock input:
• NEGative
The falling edge starts the period in which the input data is sampled.
• POSitive
The rising edge starts the period in which the input data is sampled.
This command has restrictions for frequencies under 620Mbits/s. See for details.
See the User Guide (or online Help) for details.
SOURce7 Subsystem
SOURce7 Subsystem - Reference
SOURce7 represents the error detector's Trigger Out port.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 279
Description
IVI-COM Equivalent
Syntax
Description