Agilent Technologies J-BERT N4903B Manual de Serviço Página 351

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 406
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 350
Returns a comma-separated list of measured values for nItems data points. If fewer
points are available, only the values for available points are returned.
The list has the following parameters:
First value returned is number of data points actually returned.
Second value is a flag that indicates whether still more points are available.
For each returned measurement point, 10 values are returned:
dDelay: Delay (data type: NR3).
dThreshold: Value for the bit error rate threshold (data type: NR3).
dComparedBits: Number of compared bits (data type: NR3).
dErroneousBits: Number of errors (data type: NR3).
dErroneousZeros: Number of errors from zeros (data type: NR3).
dErroneousOnes: Number of errors from ones (data type: NR3).
bExtrapolatedFlag: Flag 0 or 1 indicating if the data was calculated or
extrapolated by the firmware server.
Delay (data type: NR3) as set by “MEAS:FEMask(*):PARameter:POINt
[:VALue][?]” on page 349
Threshold (data type: NR3) as set by “MEAS:FEMask(*):PARameter:POINt
[:VALue][?]” on page 349
ThresholdType (data type: NR3) as set by “MEAS:FEMask
(*):PARameter:TRESolution:TYPE[?]” on page 348
MEAS:FEMask(*):POINt:PASSed?
IAgilentN490xEDFastEye.GetMaskPoint
MEAS:FEMask(*):POINt:PASSed? nIndex
Returns whether the specified measurement point has passed.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 351
Description
IVI-COM Equivalent
Syntax
Description
Vista de página 350
1 2 ... 346 347 348 349 350 351 352 353 354 355 356 ... 405 406