The query returns the present setting.
SOURce8[:JITTer]:PERiodic2:SWEep:FREQuency:STOP[?]
SOUR8[:JITT]:PER2:SWE:FREQ:STOP <NR3>
SOUR8[:JITT]:PER2:SWE:FREQ:STOP?
The command defines the stop frequency of the periodic jitter sweep. The stop
frequency has to be higher than start frequency and the range should be in
accordance with the Waveform. This command works only if SOUR8
[:JITT]:PER2:SWE:LEV[:VAL] CONS is selected.
The query returns the present setting.
SOURce8[:JITTer]:PERiodic2:SWEep:TIME[?]
SOUR8[:JITT]:PER2:SWE:TIME <NR3>
SOUR8[:JITT]:PER2:SWE:TIME?
The command defines the time duration of the sweep.
The query returns the present setting.
SOURce8[:JITTer]:PERiodic2:SWEep:STEP[:VALue][?]
SOUR8[:JITT]:PER2:SWE:STEP[:VAL] <NR1>
SOUR8[:JITT]:PER2:SWE:STEP[:VAL]?
The command defines the number of steps to fulfill a complete sweep. The start
and stop values are included. The valid range is from 2 to 100.
The query returns the present setting.
SOURce8[:JITTer]:PERiodic2:SWEep:STEP:DISTance[?]
SOUR8[:JITT]:PER2:SWE:STEP:DIST EQU | AUTO
SOUR8[:JITT]:PER2:SWE:STEP:DIST?
SCPI Command Reference 5