Vista de página 316
Table 90
Name Description under
STATus:PROTection STATus:PROTection Subnode” on page
325
STATus:SYMBol “STATus:SYMBol Subnode” on page
327
STATus:PRESet
IAgilentN490xStatus.Preset (not IVI-compliant)
STATus:PRESet
The PRESet command is an event that configures the SCPI and device-dependent
status data structures, such that the device-dependent events are reported at a
higher level through the mandatory part of the status reporting structures.
The PRESet command affects only the enable register and the transition filter
registers for the SCPI mandated and device dependent status data structures.
PRESet does not affect either the "status byte" or the "standard event status" as
defined by IEEE 488.2. PRESet does not clear any of the event registers. The *CLS
command is used to clear all event registers in the device status reporting
mechanism.
From the device-dependent status data structures, the PRESet command sets the
enable register to all one's and the transition filter to recognize both positive and
negative transitions. For the SCPI mandatory status data structures, the PRESet
command sets the transition filter registers to recognize only positive transitions
and sets the enable register to zero.
SCPI Command Reference 5