Operating the 8000 4
8000 Operator’s Manual 91
The following conventions are used with this screen:
• Review previously entered sample time points by pressing ENTER
to scroll through the entries.
• After all desired sample time points have been entered, enter 000:00
for the last sample time point and press ENTER. If the maximum of
ten sample time points are entered, you do not have to enter 000:00
for the last sample time point.
• To add additional sample time points, press ENTER until a new row
time displays. Enter the additional sample time point.
• If more than ten sample time points are needed, use the link program
feature. See “Starting a Program” on page 91.
• To correct an entry, press BACKSPACE to go to the previous
character position and re-enter the correct value.
• If a shorter sample time point than the previous one is entered for a
given row, the row number will not increment. Times must increase
from one row to the next.
• To shorten a saved program, enter 000:00 for the first sample time
point you wish to discard. All successive time points are eliminated.
• Press ESC to save all new values and return to the Program Variables
screen.
Starting a Program
1 From the Ready screen, press START PROG to start a program or
modify a program already operating.
2 Enter a program number between 1 and 15 and press ENTER. The
Link Program Option screen displays.
Sample time points should be at least five minutes apart to allow
adequate time for priming and purging the sampling lines. If your
system configuration includes the syringe pump and filter changer,
ensure the sample time points are at least six minutes apart.
***LINK PROGRAM OPTION***
1 RUN SINGLE PROG 2 LINK ANOTHER PROG