Operating the 8000 4
8000 Operator’s Manual 65
Program Option 2 - Sample Time Points
From the Program Variables screen, select option 2, Sample Time
Points. The following screen displays:
The following conventions are used with this screen:
• Review previously entered sample time points by pressing ENTER
to scroll through the entries.
• After all desired sample time points have been entered, enter 000:00
for the last sample time point and press ENTER. If the maximum of
ten sample time points are entered, you do not have to enter 000:00
for the last sample time point.
• To add additional sample time points, press ENTER until a new row
time displays. Enter the additional sample time point.
• If more than ten sample time points are needed, use the link program
feature. See “Starting a 8000 Program” on page 66.
• To correct an entry, press BACKSPACE to go to the previous
character position and re-enter the correct value.
SAMPLE TIME POINT: 000:02
ENTER hhh:mm UP TO 999:59 HOURS
PROG #: 1 ROW: 0 VOLUME: 5ml
TIME:01/01/12 09:37:14 ELAPSE:000:00:00
This screen accepts sampling times and saves them for execution when
the program is run. Time points are requested in order of execution
starting with collection Row 0 on the sample tray. The sample time is
entered as hours and minutes. Once you enter a time, the row number
increments and another sample time point is requested.
If a sample time point is required after the final spin (250 RPM), enter a
maximum of nine sample time points. The firmware accounts for the
final spin sample, but the tenth space must be left blank.