76 8000 Operator’s Manual
4 Operating the 8000
4 Manual Prime Vol.
(syringe pump)
If your system configuration includes the syringe pump, use this option to set the
amount of drawn medium necessary to fill the sampling lines of the entire system.
Select option 4, Manual Prime Vol. Enter a value for the priming volume and press
ENTER. The Main Menu displays.
Note: At a minimum, the manual prime volume should be large enough for medium
to be seen dripping from the return cannulas. In general, the longer the tubing
lengths, the larger the priming volume required. This value is not used by sampling
programs which have their own individual priming volume values. It is used only
with the MANUAL SAMPLE and CLEAN SYSTEM keys on the 8000.
5 Manual Purge Time
(peristaltic pump)
If your system configuration includes the peristaltic pump, use this option to set the
interval the peristaltic pump runs in reverse to return the uncollected medium in each
line back into the vessel from which it was drawn.
Select option 5, Manual Purge Time. Enter a value for the manual purge time from 1
to 99 seconds and press ENTER. The Main Menu displays.
Note: Purging also clears the sampling filters of particulate matter that may restrict
successive samples. As with priming, the longer the tubing lengths, the longer the
purge time required. This value is not used by sampling programs, which have their
own individual purging time values. It is used only with the MANUAL SAMPLE,
CAL, and CLEAN SYSTEM keys on the 8000.
5 Manual Purge Vol.
(syringe pump)
If your system configuration includes the syringe pump, use this option to set the a
purge volume that ensures all stranded medium is properly expelled.
Select option 5, Manual Purge Vol. Enter a value for the purge volume that ensures at
least two strokes of the plunger and press ENTER. The Main Menu displays.
Note: The first stroke moves all medium from the sampling lines to the return lines.
The second stroke draws air into the sampling lines and purges the remaining
medium from the return lines. As with priming, the longer the tubing lengths, the
larger the purge volume required. This value is not used by sampling programs
which have their own individual purge volume values. It is used only with the
MANUAL SAMPLE and CLEAN SYSTEM keys on the 8000.
6 Control APP 3/7 Use this option to remotely control the Apparatus 3 / 7 from the 8000. See “Main
Menu Option 6 - Control APP 3/7” on page 82.
7 Manual Operation Use this option to manually control the 8000. See “Main Menu Option 7 - Manual
Operation” on page 83.