Supplemental Operator’s Guide
for the Agilent 35670A Dynamic Signal Analyzer
By: David Forrest
Seattle Sound and Vibration, inc.
Print Date: December, 2000
ãSeattle Sound and Vibration, inc. 1996, 2000.
All rights reserved.
Agilent Part number 35670-90059
Special Permission Amoco Corporation and Agilent Technologies, Inc are granted
full editing, duplication and distribution privileges.
Vista de página 0
1 2 3 4 5 6 ... 81 82

Resumo do Conteúdo

Página 1 - Supplemental Operator’s Guide

Supplemental Operator’s Guidefor the Agilent 35670A Dynamic Signal AnalyzerBy: David ForrestSeattle Sound and Vibration, inc.Print Date: December, 200

Página 2

AGILENT 35670ASupplemental Operator’s Guide6 Any frequency below 51.2 kHz may of course be entered, however the analyzer willselect the lowest span th

Página 3

AGILENT 35670ASupplemental Operator’s Guide7Input CH1 Rng: 501.7819 mVpk CH2 Rng: 631.706 mVpk CH3 Rng: 2.5149 Vpk CH4 Rng: 2.5149 Vpk Date:

Página 4

AGILENT 35670ASupplemental Operator’s Guide8Section 5 : Improving Measurement ResultsFrequency Zooming to Increase ResolutionThe AGILENT 35670A has th

Página 5

AGILENT 35670ASupplemental Operator’s Guide9 If continuous real-time measurements over a long duration are required, andabsolutely no overload conditi

Página 6

AGILENT 35670ASupplemental Operator’s Guide10• Systm Util Date: 10-02-96 Time: 08:40 PM Date: 01-02-04 Time: 12:44:00 PMA: CH1 Pwr Sp

Página 7

AGILENT 35670ASupplemental Operator’s Guide11Section 6 : Measuring a Single Channel Time WaveformThis section contains procedures for setting up and m

Página 8

AGILENT 35670ASupplemental Operator’s Guide12Selecting Time Record LengthThe default time record length at a frequency span of 51.2 kHz is only 7.8 ms

Página 9

AGILENT 35670ASupplemental Operator’s Guide13• Press the [ARM] softkey (F9) again.Event TriggeringThe AGILENT 35670A has the ability to trigger the ac

Página 10 - • Press the [Scale] hardkey

AGILENT 35670ASupplemental Operator’s Guide14Input CH1 Rng: 1.0012 Vpk CH2 Rng: 631.706 mVpk CH3 Rng: 2.5149 Vpk CH4 Rng: 2.5149 Vpk Date:

Página 11 - • Press the [Marker] hardkey

AGILENT 35670ASupplemental Operator’s Guide15Section 7 : Measuring Dual Channel SpectraThis section contains procedures for setting up and measuring a

Página 12

AGILENT 35670ASupplemental Operator’s GuideSECTION 1 : TURNING ON THE AGILENT 35670A ...

Página 13

AGILENT 35670ASupplemental Operator’s Guide16• Press the [Meas Data] hardkey.• Press the [CHANNEL 1 2 34] softkey (F1) until 2 is highlighted.• Press

Página 14 - AGILENT 35670A

AGILENT 35670ASupplemental Operator’s Guide17 Use the marker knob to read the amplitude at any measured frequency for Channel 1in Trace A. To set up

Página 15

AGILENT 35670ASupplemental Operator’s Guide18Section 8 : Improving Dual Channel Measurement ResultsMaximizing Dynamic RangeAutoranging to maximize the

Página 16

AGILENT 35670ASupplemental Operator’s Guide19Avg[FFT]Type: RMS Number: 10 Update Rt: 5 Overlap: 0 % Dat

Página 17

AGILENT 35670ASupplemental Operator’s Guide20Section 9 : Measuring Dual Channel Time WaveformsThis section contains procedures for setting up and meas

Página 18

AGILENT 35670ASupplemental Operator’s Guide21• Press the [CHANNEL 1 234] softkey (F1) until 1 is highlighted.• Press the [TIME CHANNEL 1] softkey (F5)

Página 19

AGILENT 35670ASupplemental Operator’s Guide22 Then set up trigger parameters (for a positive event, positive slope) as follows:• Press the [TRIGGER SE

Página 20

AGILENT 35670ASupplemental Operator’s Guide23Using Markers with Dual Channel Time MeasurementsCompare the time trace of Channel 1 with Channel 2 using

Página 21

AGILENT 35670ASupplemental Operator’s Guide24• Press the [Active Trace] hardkey.• Press the [ABCD] softkey (F7).• Press the [Scale] hardkey.• Toggle t

Página 22

AGILENT 35670ASupplemental Operator’s Guide25Section 11 : Modal Testing Using a Hammer and AccelerometerHammer testing is often the easiest and quicke

Página 23

AGILENT 35670ASupplemental Operator’s GuideSECTION 11 : MODAL TESTING USING A HAMMER AND ACCELEROMETER ... 25SETTING UP

Página 24

AGILENT 35670ASupplemental Operator’s Guide26• Toggle the [XDCR UNIT ON OFF] softkey (F4) until ON is highlighted.• Press the [XDCR UNIT LABEL] softke

Página 25

AGILENT 35670ASupplemental Operator’s Guide27• Press the [TRIGGER SETUP] softkey (F6)• Press the [ALL CHANNELS] softkey (F8)• Press the [CHANNEL * DEL

Página 26 - Trigger Delay

AGILENT 35670ASupplemental Operator’s Guide28Window[FFT]Type: Force/Expo Width: 15 ms Decay: 30 ms Date: 10-02-96 Time: 09

Página 27

AGILENT 35670ASupplemental Operator’s Guide29window time constant of 62.5 ms. To set up these Force/Response windows, followthese steps:• Press the [

Página 28

AGILENT 35670ASupplemental Operator’s Guide30significantly. Ten averages may take longer, but may be necessary when testingdifficult, non-linear stru

Página 29

AGILENT 35670ASupplemental Operator’s Guide31Changing Frequency SpanChanging frequency span or resolution affects three important parameters in hammer

Página 30 - Specifying Trigger Parameters

AGILENT 35670ASupplemental Operator’s Guide32Section 12 : Hammer Test Setup Without a Force TransducerHammer testing for modal analysis without a forc

Página 31 - WAITING FOR CH 1 TRIGGER

AGILENT 35670ASupplemental Operator’s Guide33Section 13 : Order Domain Results in List ModeMachinery run-ups and run-downs can be analyzed with order

Página 32 - Using Force/Response Windows

AGILENT 35670ASupplemental Operator’s Guide34• Enter 4 and press the [ORDERS] softkey (F1).• Enter 5 and press the [ORDERS] softkey (F1).• Enter 6

Página 33 - Averaging

AGILENT 35670ASupplemental Operator’s Guide35Section 14 : Comparing Two-Channel Real-Time Spectra with Recalled DataCompare two channels of real-time

Página 34

AGILENT 35670ASupplemental Operator’s GuideSECTION 21 : USING AN EXTERNAL TRIGGER FOR TIME AVERAGING ... 60SPE

Página 35 - Changing Frequency Span

AGILENT 35670ASupplemental Operator’s Guide36• Press the [MORE CHOICES] softkey (F9).• Press the [DATA REGISTER] softkey (F8).• Press the [D2] softkey

Página 36

AGILENT 35670ASupplemental Operator’s Guide37• Press the [Save/Recall] hardkey.• Press the [RECALL DATA] softkey (F4).• Press the [RCL TRACE AND SCALE

Página 37

AGILENT 35670ASupplemental Operator’s Guide38 Scaling for traces A and B are now identical. To accomplish the same trace scalingfor C and D, follow t

Página 38 - To start taking measurements

AGILENT 35670ASupplemental Operator’s Guide39Section 15 : Waterfall Spectra at Time IntervalsWaterfalls on the AGILENT 35670A can be used to examine a

Página 39

AGILENT 35670ASupplemental Operator’s Guide40• Press the [Disp Format] hardkey.• Press the [WATERFALL SETUP] softkey (F8).• Toggle the [WATERFALL ON O

Página 40

AGILENT 35670ASupplemental Operator’s Guide41 This makes the waterfall trace active.• Press the [Marker Fctn] hardkey.• Press the [WATERFALL MARKERS]

Página 41 - Scaling the Displays

AGILENT 35670ASupplemental Operator’s Guide42Section 16 : Waterfall Spectra at RPM IntervalsSpectra can be acquired at fixed RPM intervals with RPM ST

Página 42 - Using Markers to Compare

AGILENT 35670ASupplemental Operator’s Guide43• Press 50 then the [RPM] softkey (F1). Then specify the total number of spectra to be collected (15):•

Página 43

AGILENT 35670ASupplemental Operator’s Guide44Marker Fcn[Watrfall]Trace: B Save Reg: D2 Base Supr: 0 % Height: 57 %

Página 44 - Setting Up Time Step Arm

AGILENT 35670ASupplemental Operator’s Guide45Section 17 : Two-Channel Absolute and Differential Amplitude MeasurementMeasuring the relative motion bet

Página 45

AGILENT 35670ASupplemental Operator’s Guide1Section 1 : Turning On the AGILENT 35670AThe entire turn-on procedure for the AGILENT 35670A takes 40 seco

Página 46 - Specifying the Start RPM

AGILENT 35670ASupplemental Operator’s Guide46• Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted..• Press the [Active Trace] hardkey.

Página 47 - Scaling the Display

AGILENT 35670ASupplemental Operator’s Guide47• difference in motion between Channel 1 and Channel 2 in mils peak-to-peak• phase angle between the tw

Página 48

AGILENT 35670ASupplemental Operator’s Guide48• Toggle the [CHANNEL 1 2 34] softkey (F1) until 2 is highlighted.• Press the [LIN SPEC CHANNEL 2] softk

Página 49 - Accelerometer Polarity

AGILENT 35670ASupplemental Operator’s Guide49• Press the [Active Trace] hardkey again.• Press the [D] softkey (F2).• Press the [Trace Coord] hardkey.•

Página 50

AGILENT 35670ASupplemental Operator’s Guide50Section 18 : Measuring Frequency Response with the AGILENT 35670AFrequency response functions characteriz

Página 51

AGILENT 35670ASupplemental Operator’s Guide51• Attach one end of stinger to shaker and one end to force transducer• Connect amplifier output to shak

Página 52

AGILENT 35670ASupplemental Operator’s Guide52 Choose measurement parameters• Press the [Freq] hardkey• Enter 800 and then press the [Hz] softkey (F2)

Página 53 - • Press the [D] softkey (F2)

AGILENT 35670ASupplemental Operator’s Guide53• Turn up the level control on the shaker amplifier until the excitation isnoticeable, but does not buzz

Página 54

AGILENT 35670ASupplemental Operator’s Guide54• Press the [Disp Format] hardkey.• Press the [SINGLE] softkey (F1)• Press the [Active Trace] hardkey.• P

Página 55

AGILENT 35670ASupplemental Operator’s Guide55• Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.• Press the [Active Trace] hardkey.•

Página 56

AGILENT 35670ASupplemental Operator’s Guide2Section 2 : Measurement State After Turn-onInst Mode FFT Channels 2Measurement Source INPUTMeas Data Data

Página 57

AGILENT 35670ASupplemental Operator’s Guide56Section 19 : Two Spectral Traces Showing Mils and Ips while EU is GMeasure an accelerometer signal and di

Página 58

AGILENT 35670ASupplemental Operator’s Guide57• Press the [Trace Coord] hardkey.• Press the [Y UNITS] softkey (F8)• Toggle the [AMPLITUDE PK PP RMS] so

Página 59 - Assessing Measurement Quality

AGILENT 35670ASupplemental Operator’s Guide58Section 20 : Peak Hold During a Machine Run-up and Coast-DownPeak amplitudes experienced during run-ups a

Página 60

AGILENT 35670ASupplemental Operator’s Guide59• Press the [Active Trace] hardkey.• Press the [A] softkey (F5).• Press the [Scale] hardkey.• Toggle the

Página 61

AGILENT 35670ASupplemental Operator’s Guide60Section 21 : Using an External Trigger for Time AveragingUse external triggering for synchronous averagin

Página 62

AGILENT 35670ASupplemental Operator’s Guide61 Acquire and Display the Trigger Signal• Press the yellow [Start] hardkey.• Press the [Freq] hardkey.•

Página 63

AGILENT 35670ASupplemental Operator’s Guide62Measuring Time Averaged Spectrums with External TriggeringNow that the external trigger has been set up,

Página 64

AGILENT 35670ASupplemental Operator’s Guide63Section 22 : Saving Trace Data to a 3.5” Disk or Non-Volatile RAMSave trace data into non-volatile RAM or

Página 65 - Setting Up External Trigger

AGILENT 35670ASupplemental Operator’s Guide64Saving Trace to Non-Volatile RAM (NV-RAM)To store Trace A data as a file named “N1.DAT” into NV-RAM, foll

Página 66

AGILENT 35670ASupplemental Operator’s Guide65Section 23 : Recalling Trace Data From 3.5” Disk or Non-Volatile MemoryRecall trace data from a file in n

Página 67 - Saving Trace to 3.5” Disk

AGILENT 35670ASupplemental Operator’s Guide3Section 3 : Measuring with TransducersThe AGILENT 35670A can be used with a broad range of transducers. T

Página 68 - Confirming Contents of NV-RAM

AGILENT 35670ASupplemental Operator’s Guide66• Press the [DATA REGISTER] softkey (F8).• Press the [D2] softkey (F2).• Press the [Save/Recall] hardkey.

Página 69 - Recall Trace From 3.5” Disk

AGILENT 35670ASupplemental Operator’s Guide67Section 24 : Plotting and Printing Trace DataUse the plotting and printing capability of the AGILENT 3567

Página 70

AGILENT 35670ASupplemental Operator’s Guide68• Select the [SYSTEM CONTROLLR] softkey (F2). Check the Plot/Print Destination• Press the [Plot/Print] ha

Página 71 - Plotting the Display

AGILENT 35670ASupplemental Operator’s Guide69Section 25 : Importing Plots into Microsoft WordThe plotter output of the AGILENT 35670A conforms to the

Página 72 - Printing the Display

AGILENT 35670ASupplemental Operator’s Guide70Plot to a File Using the AGILENT 35670AThe AGILENT 35670A has the ability to route plotter output to a fi

Página 73

AGILENT 35670ASupplemental Operator’s Guide71Section 26 : Running a Single Calibration Test on CommandThe AGILENT 35670A automatically runs periodic s

Página 74

AGILENT 35670ASupplemental Operator’s Guide72Section 29 : Transducer Unit Conversion with the AGILENT 35670AAutomatic conversion between acceleration,

Página 75

AGILENT 35670ASupplemental Operator’s Guide73• Press the [XDCR UNIT CONVERT] softkey (F4)• Press the [inch (DISP)] softkey (F7)• Press the [Scale] har

Página 76

AGILENT 35670ASupplemental Operator’s Guide74• Press the [DEFINE CONSTANT] softkey (F2).• Press the [DEFINE K1] softkey (F1).• Enter 386.4 and press

Página 77

AGILENT 35670ASupplemental Operator’s Guide75Section 30 : Using the Agilent Technologies Demostration Unit with the AGILENT35670ASeveral of the sectio

Página 78

AGILENT 35670ASupplemental Operator’s Guide4For example, if an accelerometer has am internal sensitivity of 25 pC/g, and a chargeamplifier has a gain

Página 79

AGILENT 35670ASupplemental Operator’s Guide76above the track number.The following descriptions link the Section in this User’s Guide to the appropriat

Página 80

AGILENT 35670ASupplemental Operator’s Guide77Section 4: Measuring a Single Channel SpectrumCD Track: 24CD Track Desc.: 160 kHz Calibration ToneLeft Ch

Página 81

AGILENT 35670ASupplemental Operator’s Guide78Section 10: Measuring Dual Channel Spectra and Time WaveformsCD Track: 41CD Track Desc.: Current Probe, T

Página 82

AGILENT 35670ASupplemental Operator’s Guide5Section 4 : Measuring a Single Channel Power Spectrum - DisplacementThis section contains procedures for s

Comentários a estes Manuais

Sem comentários