Operating the 8000 4
8000 Operator’s Manual 55
Once all values have been entered, press ESC. A confirmation screen
displays. Press Y to send the newly entered configuration to the
dissolution apparatus and override the existing settings.
Main Menu Option 3 - Set Syringe Pump
This section is applicable only to systems including a syringe pump.
From the Main Menu, select option 3, Set Syringe Pump. Select option
1 to enable the syringe pump. Once enabled, the Set Syringe Pump
Parameters screen displays.
Following is a description of the Set Syringe Pump Parameters screen
options:
NOTE
Ensure the syringe pump is enabled prior to setting the syringe pump
parameters (see MENU > 0 > P under “Hidden Key Functions” on page
53).
***SET SYRINGE PUMP PARAMETERS***
1 SYRINGE SIZE 2 PLUNGER SPEED
3 PRIME LOSS VOLUME 4 ASPIRT. DWELL TIME
Table 6 Set Syringe Pump Parameters (Apparatus 1 / 2)
Option Function
1 Syringe Size Use this option to set the syringe size.
Select option 1, Syringe Size. Select the syringe size that corresponds with the
Kloehn syringes loaded into the syringe pump. The Set Syringe Pump Parameters
screen displays.
2 Plunger Speed Use this option to configure the syringe pump plunger to operate at a specific
speed.
Select option 2, Plunger Speed. Enter a plunger speed between 100 and 1500
motor steps per second (sps) and press ENTER. The recommended value is 500.
The Set Syringe Pump Parameters screen displays.