Maintenance and Troubleshooting 5
8000 Operator’s Manual 111
the peristaltic pump reverses and the cleaning solution is pulled back
through the sample lines
the pump reverses again, the valves open and the cleaning solution is
expelled into the rinse chamber
the 8000 activates the valves several times to clean the valves and
needles.
Clean System Function with Syringe Pump
Apparatus 1 / 2
1 Place the rinse tray fitted with cleaning solution, usually water, on
the vessel table of the dissolution apparatus.
2 Immerse all six or eight sampling cannulas.
3 Press CLEAN SYSTEM.
4 The 8000 immediately starts to clean the system.
Apparatus 3 / 7
To clean the system, immerse all sampling cannulas into a cleaning
solution, usually water. Complete the following steps:
1 Fill one row of outer media tubes with the appropriate cleaning
solution, usually water.
2 On the Apparatus 3 / 7, press START DIP.
3 Select option 3, Goto Row.
4 Enter the row number which contains the cleaning solution and press
ENTER. The drive unit moves to the selected row.
5 Select option 2, Set Speed (DPM).
6 Enter 5 DPM and press ENTER.
7 Select option 1, Start. The agitator shafts lower and begin dipping.
8 On the 8000, press CLEAN SYSTEM. The Cleaning Systems
Selections screen displays and the cleaning procedure begins
immediately.