33-3
RF Generator Subsystem
AMPLitude1
AMPLitude1
Description Sets/queries the amplitude of the RF Generator when the RF IN/OUT port is selected.
GPIB and display units are dBm, Volts (V) and Watts (W); Default GPIB and display unit
is dBm.
Syntax RFGenerator:AMPLitude1?
RFGenerator:AMPLitude1 <real> | [:FNUM]
Options Refer to Appendix B.
AMPLitude1:ATTenuation[:AUTO]
Description Selects the ATTenuation of the RF IN/OUT port automatically each time a change of
amplitude setting occurs when ON.
Syntax RFGenerator:AMPLitude1:ATTenuation[:AUTO]?
RFGenerator:AMPLitude1:ATTenuation[:AUTO] <string>
Options ‘ON’ | ‘OFF’
AMPLitude2
Description Sets/queries the amplitude of the RF Generator when the AUX RFOUT port is selected.
GPIB and display units are dBm, Volts (V) and Watts (W); Default GPIB and display unit
is dBm.
Syntax RFGenerator:AMPLitude2?
RFGenerator:AMPLitude2 <real> | [:FNUM]
Options Refer to Appendix B.
AMPLitude2:ATTenuation[:AUTO]
Description Selects the ATTenuation of the AUX RFOUT port automatically each time a change of
amplitude setting occurs when ON.
Syntax RFGenerator:AMPLitude2:ATTenuation[:AUTO]?
RFGenerator:AMPLitude2:ATTenuation[:AUTO] <string>
Options ‘ON’ | ‘OFF’