14-3
DSP Analyzer Subsystem
AMPLitude:MARKer:POSition:FALL
AMPLitude:MARKer:POSition:FALL
Description Sets/queries the AMPLitude MARKer FALL trace position setting. The value is given in
units of divisions from the left side of the FALL trace
(144 Bit Periods (T) to 156 Bit Periods (T) = 6 divisions).
Syntax DSPanalyzer:AMPLitude:MARKer:POSition:FALL?
DSPanalyzer:AMPLitude:MARKer:POSition:FALL <real> | [:FNUM]
Options Refer to Appendix B.
AMPLitude:MARKer:POSition:MID
Description Sets/queries the AMPLitude MARKer MID trace position setting. The value is given in
units of divisions from the left side of the MID trace
(-10 Bit Periods (T) to 160 Bit Periods (T) = 8.5 divisions).
Syntax DSPanalyzer:AMPLitude:MARKer:POSition:MID?
DSPanalyzer:AMPLitude:MARKer:POSition:MID <real> | [:FNUM]
Options Refer to Appendix B.
AMPLitude:MARKer:POSition:RISE
Description Sets/queries the AMPLitude MARKer RISE trace position setting. The value is given in
units of divisions from the left side of the RISE trace
(-8 Bit Periods (T) to 4 Bit Periods (T) = 6 divisions).
Syntax DSPanalyzer:AMPLitude:MARKer:POSition:RISE?
DSPanalyzer:AMPLitude:MARKer:POSition:RISE <real> | [:FNUM]
Options Refer to Appendix B.
AMPLitude:PMZero
Description Zeroes the Power Meter in order to make calibrated Average TX Power measurements.
Note: The user should disconnect the input signal when selecting this. This field is the same
as CW:PMZero.
Syntax DSPanalyzer:AMPLitude:PMZero
Options Not Applicable.