15-12
DSP Analyzer Commands (Measure Subsystem)
PHASe:MARKer:ERRor
PHASe:MARKer:ERRor
Description Queries the PHASe ERRor measurement result. This is the y-axis MARKer position of the
phase error. This value is a function of the marker position set or queried by
DSP:PHAS:MARK:POS.
IMPORTANT: The user MUST be on the Phase Err screen to query this result
(DISP:DSP:VIEW ’PHASE ERR’).
GPIB unit is degrees.
Display unit is degrees.
Syntax MEASure:DSPanalyzer:PHASe:MARKer:ERRor[:MM] | [:AVG]
Options Refer to Appendices D and F.
PHASe:MARKer:TIME
Description Queries the MARKer TIMe which is the marker’s position relative to bit zero in the
measured burst. This value is a function of the marker position set or queried by
DSP:PHAS:MARK:POS.
GPIB units are seconds (S), bit periods (T);
default unit is seconds (S).
Display units are US (micro-second), T (bit periods);
default unit is US (micro-second).
IMPORTANT: The user MUST be on the Phase Err screen to query this result
(DISP:DSP:VIEW ’PHASE ERR’).
Syntax MEASure:DSPanalyzer:PHASe:MARKer:TIME?
MEASure:DSPanalyzer:PHASe:MARKer:TIME[:MM] | [:AVG]
Options Refer to Appendices D and F.
PHASe:MBURst
Description Selects/queries state of multi-burst measurement.
Syntax PHASe:MBURst?
MEASure:DSPanalyzer:PHASe:MBURst <string>
Options ‘ON’ | ‘OFF’