Troubleshooting and Test Functions
Initialization Failed
Initialization Failed
The quaternary pump failed to initialize successfully within the maximum
time window.
A maximum time is assigned for the complete pump-initialization cycle. If the
time is exceeded before initialization is complete, the error message is
generated.
Probable Causes Blocked active inlet valve.
Defective pump drive assembly.
Defective LPM board.
Suggested Actions Exchange the active inlet valve.
Exchange the pump drive assembly.
Exchange the LPM board.