Agilent Technologies 1100 Series Manual de Instruções Página 111

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 272
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 110
111
Repairing the Pump
Removing and Disassembling the Pump Head Assembly
3 Place the pump head on a flat surface. Loosen
the lock screw (two revolutions) and while
holding the lower half of the assembly
carefully pull the pump head away from the
plunger housing.
4 Remove the support rings from the plunger
housing and lift the housing away from the
plungers.
Pump head
Lock screw
Plunger housing
Support ring
Plunger housing
Plunger
Vista de página 110
1 2 ... 106 107 108 109 110 111 112 113 114 115 116 ... 271 272

Comentários a estes Manuais

Sem comentários