Troubleshooting and Test Functions
Valve Fuse
Valve Fuse
Valve Fuse 0: Channels A and B
Valve Fuse 1: Channels C and D
The gradient valve in the quaternary pump has drawn excessive current
causing the electronic fuse to open.
Probable Causes Defective gradient valve.
Defective connection cable (front panel to LPM board).
Defective LPM board.
Suggested Actions Restart the quaternary pump. If the error message appears again, exchange
the gradient valve.
Exchange the connection cable.
Exchange the LPM board.