Control Module Screens for the Quaternary Pump
Screens available from the Records screen
Info Log Use the m-key to receive a pop-up menu, Select Info Log. A list of the last
events are listed. For troubleshooting reasons they can be printed or saved to
a file on the PCMCIA card (using the m-key for the context sensitive menu).
EMF (Early
Maintenance
Feedback)
Use the F1 key (EMF) to set EMF parameters. Choose menu item 1 (Setup
limits) to select the amount of pumped solvent or seal wear, when you want
to receive a warning for exceeded limits. Press F7 (Reset) to reset the
counters after exchanging parts that are subject to wear.