7
Trigger Function
Trigger type Continuous, Single, Number of groups
Trigger source Internal (Free run), External (BNC connector input), GPIB or LAN,
Manual (Front key)
Trigger event type Point trigger, Sweep trigger
Measurement Bandwidth/Averaging
Measurement bandwidth
Range 1 (Fast) to 5 (Precise), 5 steps
Averaging
Type Sweep-to-sweep averaging, Point averaging
Averaging factor 1 to 256 (integer)
Adapter Setup
Adapter Selection
NONE No adapter (the 16047E, etc. direct connection type test fixture is
connected)
4TP 1M Four-terminal pair 1 m (16048G)
4TP 2M Four-terminal pair 2 m (16048H)
7-mm 42942A Terminal Adapter (42942A)
PROBE 42941A Impedance Probe (42941A)
Calibration
Calibration
User calibration Calibration performed with user-defined calibration kit (OPEN,
SHORT, LOAD)
Port extension Compensation performed when the measurement terminal is
expanded from the 7-mm connector of the 42942A Terminal
Adapter or the 3.5-mm connector of the 42941A Impedance
Probe. Enter electrical length or delay time for the extension.
Fixture compensation Compensation performed at the device contacts of the test fix-
ture using OPEN, SHORT, LOAD.
Calibration points Fixed points, or User points determined by sweep setups